| Model Number |
26050CA |
| Medical Device Problem Codes |
Material Separation (1562); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a laser auriga 30 fragmentation of a large bladder stone with the storz resector, the ceramic tip broke off and passed into the bladder.This forced the surgeon to do a laparotomy to recover the part inside the lava.The ceramic completely separated from the sheath and passed into the bladder.The surgery was prolonged for 1h 30 minutes.The resection of the prostate was rescheduled 15 days later.
|
| |
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned on 2023-09-07.The investigation of the product was completed on 2023-11-29.The evaluation of the product revealed that the articles in question were repaired by a third party.It is therefore very likely that the external company did not fulfill karl storz's quality requirements.At the distal end, the shaft of two articles was most likely machined too much, so that the wall of the tube was too thin and the connection between the ceramic beak and the tube was no longer given.On the third shank there is no recognizable glue residue, which has resulted in the ceramic beak having no connection to the shank.The missing connection between the ceramic beak and the shank, caused the ceramic to fall off.As described in the ifu that the articles needs to be checked for intactness before use, these damages should have been noticed, therefore, apart from the external repair, a user error is not excluded.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|