| Catalog Number |
80400 |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Per journal article "posttransfusion sepsis attributable to bacterial contamination in platelet collection set manufacturing facility, united states" by ian kracalik, et all.During (b)(6) 2022, we reviewed transfusion-transmitted sepsis cases in the united states attributable to polymicrobial contaminated apheresis platelet components, including acinetobacter calcoaceticus baumannii complex or staphylococcus saprophyticus isolated from patients and components.Transfused platelet components underwent bacterial risk control strategies (primary culture, pathogen reduction or primary culture, and secondary rapid test) before transfusion.Environmental samples were collected from a platelet collection set manufacturing facility.Seven sepsis cases from 6 platelet donations from 6 different donors were identified in patients from 6 states; 3 patients died.Cultures identified acinetobacter calcoaceticus baumannii complex in 6 patients and 6 transfused platelets, s.Saprophyticus in 4 patients and 4 transfused platelets.Whole-genome sequencing showed environmental isolates from the manufacturer were closely related genetically to patient and platelet isolates, indicating the manufacturer was the most probable source of recurrent polymicrobial contamination.Clinicians should maintain awareness of possible transfusion-transmitted sepsis even when using bacterial risk control strategies.Specific details, such as patient information and outcome, were not included in the article for these events, therefore this report is being provided as a summary of the events.Patient information is unknown at this time.The collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot number and expiry information are not available at this time.Investigation is in process, a follow-up report will be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6, h.10 and h.11.Lot number and expiry information are not available at this time.It has been determined that the event submitted on this report is a duplicate of the event submitted on report 1722028-2024-00088.No further reporting will be provided as this does not represent a reportable event.
|
| |
|
Event or Problem Description
|
|
Per journal article "posttransfusion sepsis attributable to bacterial contamination in platelet collection set manufacturing facility, united states" by ian kracalik, et all.During may 2018.December 2022, we reviewed transfusion-transmitted sepsis cases in the united states attributable to polymicrobial contaminated apheresis platelet components, including acinetobacter calcoaceticus baumannii complex or staphylococcus saprophyticus isolated from patients and components.Transfused platelet components underwent bacterial risk control strategies (primary culture, pathogen reduction or primary culture, and secondary rapid test) before transfusion.Environmental samples were collected from a platelet collection set manufacturing facility.Seven sepsis cases from 6 platelet donations from 6 different donors were identified in patients from 6 states; 3 patients died.Cultures identified acinetobacter calcoaceticus baumannii complex in 6 patients and 6 transfused platelets, s.Saprophyticus in 4 patients and 4 transfused platelets.Whole-genome sequencing showed environmental isolates from the manufacturer were closely related genetically to patient and platelet isolates, indicating the manufacturer was the most probable source of recurrent polymicrobial contamination.Clinicians should maintain awareness of possible transfusion-transmitted sepsis even when using bacterial risk control strategies.Specific details, such as patient information and outcome, were not included in the article for these events, therefore this report is being provided as a summary of the events.Patient information is unknown at this time.The collection set is not available for return because it was discarded by the customer.
|
| |
|