Model Number HL-90
Medical Device Problem Codes Break (1069); No Display/Image (1183)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the temperature display was damaged.Patient involvement unknown.
 
Additional Manufacturer Narrative
Other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Evaluation codes: updated device evaluation: one device was returned for investigation.Visual inspection noted: worn float switch, corroded drain fitting, cracked tank cover, and bent prongs on line cord.Functional testing found the lcd has liquid crystal leakage, confirming the customer complaint.Root cause was attributed to a faulty printed circuit board (pcb).What caused the damage to the pcb could not be established.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the float switch, drain fitting, tank cover, and line cord.Also replaced the pcb and power switch.Performed preventative maintenance.Device passed functional testing after the completed repairs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Common Device Name
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17674136
Report Number3012307300-2023-08643
Device Sequence Number6057216
Product Code LGZ
UDI-Device Identifier50695085821098
UDI-Public50695085821098
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K911383
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90-40
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/23/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2023
Supplement Date Received by Manufacturer10/13/2023
Initial Report FDA Received Date09/01/2023
Supplement Report FDA Received Date11/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/14/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-