• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTICURETEROSCOPE1.2C 3.0TP 700 ROP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTICURETEROSCOPE1.2C 3.0TP 700 ROP Back to Search Results
Model Number FUR-9RBS
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is not during procedure.There was no report of patient harm.Fiber image failure (broken).
 
Additional Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model fnl-7rp3 is available in the usa with a 510k number k951196.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cfb (coherent fiber bundle) was broken.Based on the result, we concluded that it was caused due to the excessive force applied on the cfb (coherent fiber bundle).In addition, our technician confirmed that the insertion flexible tube crushed; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586 (image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
FIBEROPTICURETEROSCOPE1.2C 3.0TP 700 ROP
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315800
MDR Report Key17680766
Report Number9610877-2023-58447
Device Sequence Number11415201
Product Code FGB
UDI-Device Identifier04961333083633
UDI-Public04961333083633
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFUR-9RBS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2023
Initial Report FDA Received Date09/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-