| Catalog Number |
289707000 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a synthes employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2023 during routine incoming inspection of a loaner set, it was observed that the torque device was found to be out of specification.There was no known patient or hospital involvement.No additional information is available.This report is for a uniplate anterior cervical plate system cam tightener torque handle.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there were no cosmetic defects within the uniplate cam tghtnr torq hndl.A dimensional inspection for the uniplate cam tghtnr torq hndl was not performed as it is not applicable to the complaint condition.A functional test was performed using torque meter eztorq iii 150i torque analyzer cd78632, adaptor 103094368.Drawing 289707000 rev.J manufactured was used as source of specification.Torque specification for the device was 0.63 - 0.98 nm.Actual measured values range from 1.014 - 1.080 nm.The device was testing high.The complaint was able to be replicated.Refer to attachment "(b)(4) torque limiting results.Pdf".As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the uniplate cam tghtnr torq hndl would contribute to the complained device issue.The under torqueing condition can be attributed to maintenance of the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: the following source controlled drawings reflecting the current and manufactured revisions were reviewed: uniplate cam tightener torque handle dwg-2897-07-000 rev.J (current and manufactured).Dimensional inspection: n/a.A review of the receiving inspection (ri) for uniplate cam tghtnr torq hndl , was conducted identifying that lot number km952904 was released in one batch.Batch1: lot units were released on 16 mar 2023 with no discrepancies.Supplier : (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|