SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLK-4
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to olympus, that the halogen light source had a broken handle, damage to the hood on two corners, ignition with unexpected noise and deformed structure.Inspection and testing of the returned device found that the spare lamp was missing.There were no reports of patient harm associated with this event.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
 
Additional Manufacturer Narrative
The device was returned for evaluation and the customers allegation was confirmed.It was noted that the issue occurred as the fan was damaged.It was also noted that the handle and hood were broken and the feet were missing.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.It is likely that the phenomenon occurred due to a shortage of spare lamps.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Common Device Name
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17686322
Report Number3002808148-2023-09390
Device Sequence Number19491591
Product Code FAJ
UDI-Device Identifier04953170024511
UDI-Public04953170024511
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K993041
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2023
Supplement Date Received by Manufacturer12/20/2023
Initial Report FDA Received Date09/05/2023
Supplement Report FDA Received Date01/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-