| Catalog Number |
0620050010 |
| Medical Device Problem Code |
Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
Unspecified Respiratory Problem (4464)
|
| Date of Event |
08/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced subcutaneous emphysema.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker is the initial importer of this product.The legal manufacturer is world of medicine (wom).Wom has been made aware of this report event.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: subcutaneous emphysema swelling of the skin.Probable root cause: because device was not returned to oem - wom, probable root cause cannot be determined.Most of the reasons for occurrence of a subcutaneous emphysema are related to surgical technique and trocar placement.Incorrect placement of a cannula, veress needle or a trocar into subcutaneous tissue may lead to emphysema, which can lead to crepitus.A big leakage of co2 around trocar incisions may also lead to subcutaneous emphysema.Other risk factors are prolonged surgeries, the use of many access points to the abdominal cavity and an elevated abdominal pressure.In addition, the event description does not include any indication for a device malfunction, nor is subcutaneous emphysema considered a serious injury.The reported failure mode will be monitored for future reoccurrence.Stryker is the initial importer of this product.The legal manufacturer is world of medicine (wom).Wom has been made aware of this report event.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced subcutaneous emphysema.
|
| |
|