MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507510MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast construction revision procedure with a mentor memorygel breast implant 510cc gel breast prosthesis that ruptured post implantation.Rupture of the patient¿s right breast prosthesis was diagnosed via a physical exam.As a result, the patient underwent explantation on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2023, the mentor failure analysis lab received the device for evaluation.Additionally, mentor became aware of the following: - an updated explantation date ((b)(6) 2023) was reported.- the patient had both breast prostheses explanted and they were replaced with a mentor memorygel breast implant 490cc gel breast prosthesis and a mentor memorygel breast implant 510cc gel breast prosthesis.Mentor completed the analysis on the suspect medical device on the same date.Device evaluation summary: according to the information received, the patient experienced a rupture in the breast implant.The product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample determined that no damage or anomalies were observed on the breast implant.Leak testing was performed in accordance with mentor procedures and (2) two punctures were found on the anterior view of the breast implant, both measuring approximately 0.1 cm.A microscopic examination was performed on the edges of the punctures, and parallel striations were found in the whole area of the (2) two punctures.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps, or scissors to contact the device during the implantation or other surgical procedures.A second product was received ruptured (lot #6602962).No adverse events were reported for this concomitant (contralateral) device.Therefore, no further analysis is required.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|