|
It was reported that on (b)(6) 2021 a biozorb was implanted in the patient and the patient reported suffering from a hard, painful, palpable lump where the biozorb was implanted.The patient also reported suffering from excruciating pain, difficulty sleeping, limited mobility, swelling, necrosis, and seroma at the site of the biozorb device.It was reported that the biozorb device did not absorb, instead the device migrated and was pressing up against the patient¿s skin, nearly protruding through her breast which is now deformed due to her complications from the device and removal procedure.As a result of the pain and complications of the biozorb device, the patient feared the possibility of another tumor, every day until the surgical removal of biozorb, causing significant emotional distress.No initial evaluation could be performed because there was no returned sample for this complaint the sample was not returned for this complaint and there was limited patient-related information provided for this event; therefore, a thorough investigation was not able to be performed.Investigation into several of the harms reported in this complaint was performed from a clinical literature review, as well as an investigation into the design and risk of the biozorb product.The harms identified during this complaint are: migration, limited mobility, pain, seroma, device erosion through skin, device explantation, failure to resorb reported, necrosis, deformity, swelling, sleep dysfunction a review of clinical literature performed on lumpectomies and their associated complications was performed.A summary of this review concluded that, while lumpectomy is a well-established option for breast cancer treatment, it inherently carries risks of complications, most notably infections, seroma formation, chronic pain, and wound healing issues.The literature demonstrates variation in complication rates based on surgical technique, additional therapies (such as intraoperative radiation therapy (iort) or radiation), and individual patient factors, with infection rates ranging from 2% to 10% and seroma rates from 11% to 65%.Migration: although the marker introduces some device-specific risks, such as adverse reaction to device materials or device migration, the overall safety profile of biozorb appears comparable to lumpectomy alone, with many complications likely related to the surgery rather than the device.Careful postoperative management and monitoring can help mitigate these risks.Additionally, biozorb shows lower adverse event rates compared to lumpectomy alone, likely due to the longer-term monitoring in biozorb studies versus the short-term follow-up typically used in lumpectomy literature.Inflammation / swelling / lymphedema / limited mobility: biozorb (0.7%): tissue damage, including wound dehiscence and erosion, was reported in 0.7% of patients with the biozorb marker.In a more severe case, the biozorb device eroded into the nipple-areola complex, necessitating complete removal of the nipple-areola complex due to chronic inflammation.However, given that wound healing issues, tissue necrosis, and other damage are known complications of lumpectomy itself, some of these outcomes could be directly related to the surgical trauma and healing process rather than the device.Lumpectomy (13.1% of patients, delayed healing in 6.5%, and skin-flap necrosis in 1.1%, indicating that tissue damage): wound healing issues and tissue necrosis are common after lumpectomy, particularly in cases where large amounts of tissue are removed or additional procedures such as radiation are performed.Knowles et al.Reported wound infection in 13.1% of patients, delayed healing in 6.5%, and skin-flap necrosis in 1.1%, indicating that tissue damage can be a lumpectomy-related issue.Regarding lumpectomy¿s, berger et al.(berger l, grimm a, sutterlin m, et al.Major complications after intraoperative radiotherapy with low-energy x-rays in early breast cancer.Strahlenther onkol.Apr 2024;200(4):276-286.Doi:10.1007/s00066-023-02128-z) retrospectively analyzed the occurrence of severe local complications in 10 out of 408 women who underwent intraoperative radiotherapy (iort) with low-energy x-rays during breast-conserving surgery (bcs) for early breast cancer between 2002 and 2017.The complications, which required surgical intervention, included redness (80%), seroma (60%), wound infection (60%), suture dehiscence (60%), and induration (40%).Hematoma and necrosis were less common (10% each).These symptoms emerged from immediately post-operation up to 15 years later, with a median onset at 3.1 months.While most complications were managed with smaller surgical interventions, such as excision of necrotic areas or secondary sutures, more severe cases required complex flap surgery or mastectomy.The study concluded that although iort is generally safe, it carries a risk of severe complications.Preoperative counseling and vigilant follow-up are recommended to manage these risks effectively.Pain: biozorb (0.9% to 1.9%): pain or discomfort was reported in 0.9% to 1.9% of patients with the biozorb marker.In one instance, discomfort occurred due to the device being sutured to the serratus muscle but did not result in removal.In another study, discomfort in 1.1% of patients led to device removal.However, given that pain is a well-known complication following lumpectomy, particularly in cases with nerve damage or radiation therapy, some of this discomfort could be attributed to the initial surgical intervention rather than the presence of the marker.Lumpectomy (25% to 60%): chronic pain is a frequent long-term complication of lumpectomy, affecting 25% to 60% of breast cancer surgery survivors.Kwee et al.Reported a pooled prevalence of 31% for neuropathic pain following bcs.The presence of post-surgical pain, particularly neuropathic pain caused by nerve damage, could be exacerbated by or misattributed to the biozorb marker in some cases.Seroma: biozorb (0.7% to 3.3%): seroma formation rates with biozorb ranged from 0.7% to 3.3%.Kaufman et al.(2021)4 observed that 82.8% of seromas occurred within the first three months after surgery, suggesting a link to the initial surgical procedure rather than the device itself.The overall seroma rate was 2.8%.Given that seromas frequently develop in response to the space left after lumpectomy, the seroma formation observed in biozorb studies may be more related to the lumpectomy rather than the device.Lumpectomy (up to 65%): seroma is a common complication following lumpectomy, with reported rates as high as 65% in some studies, particularly when combined with intraoperative radiotherapy (iort) or axillary procedures.Sfarad et al.Reported a 65% seroma rate following iort, suggesting that the surgical removal of tissue, not the marker itself, is the primary driver of this complication.Tissue damage / delayed wound healing / necrosis / scar tissue / erosion: biozorb (0.7%): tissue damage, including wound dehiscence and erosion, was reported in 0.7% of patients with the biozorb marker.In a more severe case, the biozorb device eroded into the nipple-areola complex, necessitating complete removal of the nipple-areola complex due to chronic inflammation.However, given that wound healing issues, tissue necrosis, and other damage are known complications of lumpectomy itself, some of these outcomes could be directly related to the surgical trauma and healing process rather than the device.Lumpectomy (13.1% of patients, delayed healing in 6.5%, and skin-flap necrosis in 1.1%, indicating that tissue damage): wound healing issues and tissue necrosis are common after lumpectomy, particularly in cases where large amounts of tissue are removed or additional procedures such as radiation are performed.Knowles et al.Reported wound infection in 13.1% of patients, delayed healing in 6.5%, and skin-flap necrosis in 1.1%, indicating that tissue damage can be a lumpectomy-related issue.Device removal / explantation / additional intervention / treatment / scan required: biozorb (1.3%): device removal due to adverse events, including infection, fibromatosis, pain, and anxiety, occurred in 1.3% of patients).Some removals were related to discomfort or anxiety about the device, while others were associated with complications potentially tied to the lumpectomy procedure itself, such as infection or tissue damage.Lumpectomy (percentage is not applicable): although device removal is not applicable to lumpectomy, reoperations due to complications such as infections or positive margins are not uncommon in lumpectomy patients.Physical asymmetry / deformity / disfigurement: rahimi 2022 (rahimi a, simmons a, kim dn, et al.Preliminary results of multi-institutional phase 1 dose escalation trial using single-fraction stereotactic partial breast irradiation for early stage breast cancer.Int j radiat oncol biol phys.Mar 1 2022;112(3):663-670.Doi:10.1016/j.Ijrobp.2021.10.010) concluded that cosmetic outcomes were preserved, with no significant cosmetic detriment across any dose cohort, suggesting that single-fraction s-pbi does not negatively affect patients' physical appearance post-treatment.Both patient- and physician-reported cosmetic scores were consistently favorable over 12 and 24 months of follow-up.A review of clinical literature performed on lumpectomies and their associated complications was performed.A summary of this review concluded that, while lumpectomy is a well-established option for breast cancer treatment, it inherently carries risks of complications, most notably infections, seroma formation, chronic pain, and wound healing issues.The literature demonstrates variation in complication rates based on surgical technique, additional therapies (such as intraoperative radiation therapy (iort) or radiation), and individual patient factors, with infection rates ranging from 2% to 10% and seroma rates from 11% to 65%.Chronic pain, particularly neuropathic pain, affects nearly one-third of patients.(kwee et al.(2024)), migration can be influenced by several patient related factors which can influence the likelihood of migration after lumpectomy, such as breast density, biopsy location, and biopsy approach.Per a study by ashali jain et al.(2017), marker migration is a relatively common occurrence following stereotactic biopsies, happening in 13% of cases.Swelling is one of the expected events that could occur after surgery and/or radiation therapy regardless of the use of a biozorb device.Due to this, confirmation of the swelling during this event as a normal course of surgery vs being caused by the biozorb is not able to be done.Device explantation is based on the patient¿s physician¿s evaluation of their physical and health state and recommendation; therefore, an investigation into this intervention is not able to be performed for this complaint.Though complications are relatively common, most are manageable with appropriate postoperative care, making lumpectomy a viable option for many breast cancer patients when carefully monitored and followed up.An investigation was performed by subject matter experts (sme¿s) into several of the harms identified as being experienced by the patient related to design, risk management, and clinical factors.The results of these investigations and potential causes can be seen below; however, it is important to note that these results do not include medical input from a qualified medical professional: pain: there is a lack of migration requirements and testing to evaluate if the device can cause tissue damage due to migration.Verification of mechanical strength and simulated use is not adequate due to several opportunities related to sampling size and test methods.The crystalline proportion in the device is unknown.Crystallinity may be causing irritation and foreign body reaction considering the long period of absorption of the device.There is a lack of usability testing evaluating the user needs, so there is no assurance that the surgical methods being used for the users are adequate.There is a lack of adequate user needs to capture appropriate design inputs, and therefore there are opportunities in design outputs and v&v activities.Migration: there is a lack of migration requirements and testing to evaluate if the device can cause tissue damage due to migration.The crystalline proportion in the device is unknown.High crystallinity may be causing migration/erosion considering the long period of absorption of the device.Verification of mechanical strength and simulated use is not adequate due to several opportunities related to sampling size and test methods.There is a lack of usability testing evaluating the user needs, so there is no assurance that the surgical methods being used for the users are adequate.There is a lack of adequate user needs to capture appropriate design inputs.Based on the clinical literature review, the harms reported to be experienced by the patient in the complaint are common symptoms from surgery and lumpectomies; therefore, a confirmation of the biozorb causing the harms was not able to be done.However, based on the review of the top harms in regards to the biozorb design and risk documentation, the adverse events reported can be tracked back to a potential shared cause: lack of adequate user needs for the product.User needs define the framework for product development and lay the ground for the definition of design inputs, design outputs and later verification and validation activities.Other potential causes are related to inadequate test methods and/or the lack of evidence to support some specific product requirements.Conclusion: when comparing the safety outcomes of the biozorb marker with general lumpectomy complications, it becomes evident that some of the complications associated with biozorb may stem from the lumpectomy procedure itself.Both procedures share overlapping risks, although the biozorb device introduces additional considerations.However, it has been confirmed that the lack of defined user needs and inadequate test methods and testing evidence are potential root causes for the harms reported in this complaint.This cause will be further investigated and addressed in a capa.The risks associated with this complaint event are further described and evaluated against the product and its hazard analysis in the health risk assessment biozorb complaint analysis.Updates to the biozorb hazard analysis risk file will be monitored.Future events will be monitored and trended.Complaint confirmed: no.Device history record (dhr) review: a dhr was reviewed with the corresponding lot, and the device was released meeting all qa specifications.
|