| Model Number |
HYF-XP |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported to olympus, during reprocessing, the oes hysterofiberscope tested positive on (b)(6) 2023, for 2 colony forming units (cfus) of an unspecified nonpathogenic microorganism, it was not specified which channels were sampled.The user did not report any contamination or any other serious deterioration in the state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has not yet been returned to olympus for evaluation, therefore the physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.The customer provided the cleaning, disinfection, and sterilization (cds) processes performed at the user facility.The customer did confirm that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The customer did not provide the specific steps taken during the cleaning sterilization and disinfection (cds) process.The investigation is ongoing, a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information provided by the third party microbiological results: updated field: h10 the hygiene microbiological investigation report indicated the channels of the scope were cultured and found <1 cfu of an unspecified microorganism.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.According to reviewing the instruction manual of hyf-xp, the reprocessing methods are described in the following chapters: chapter 6 compatible reprocessing methods and chemical agents.Chapter 7 cleaning, disinfection, and sterilization procedures.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|