• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 1688 AIM 4K AUTOCLAVABLE CAMERA HEAD, C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 1688 AIM 4K AUTOCLAVABLE CAMERA HEAD, C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 1688410105
Medical Device Problem Code Fogging (1253)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was foggy image.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
Alleged failure: moisture was seen behind the glass of the camera head.Probable root cause: cables, hdmi cables, connectors, digital board, power supply, filter / fuse, ac inlet board, main board, transition board, intermittence between transition board and ch connector software.Camera head (sensor, sensor cable)¿coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp, zoom ring) problem toggling light source connected monitors leaking poor grounding (e.G camera head connection from cable to transition board.) no/incorrect communication with light source soaking cap disconnection during sterilization electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge cybersecurity attack pendulum ch inertial measurement unit (imu) incident light from surgical scene is reflected by coupler/ch window use error.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 1688 AIM 4K AUTOCLAVABLE CAMERA HEAD, C-MOUNT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key17691634
Report Number0002936485-2023-00884
Device Sequence Number6057430
Product Code GCJ
UDI-Device Identifier07613327562545
UDI-Public07613327562545
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K212511
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1688410105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/15/2023
Supplement Date Received by Manufacturer08/15/2023
Initial Report FDA Received Date09/06/2023
Supplement Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/01/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-