|
It was reported that an intra-aortic balloon (iab) was being inserted to assist with weaning the patient from bypass following a valve procedure.During therapy, the cardiosave intra-aortic balloon pump (iabp) generated an autofill failure alarm and the iab would not inflate.A new iab was inserted but the same issue occurred.At the time of attempting to initiate therapy, it was noted that the patient had adequate electrocardiogram (ekg) signal to the iabp and fluid-filled transducer was set up.There were not any significant issues with insertion.They did not perform a manual fill at the time of initiation for either iab and they did not attempt to use another iabp.However, the technician did separate the second catheter from the helium tubing and felt helium flowing out of the helium extender tubing.It was also noted that when the second iab was removed, there was dried blood observed on the catheter and when inflation was attempted it was somewhat stuck together.Following the failure to initiate iab support, the patient was then placed on extracorporeal life support (ecls) and taken from the operating room (or) for a short period of time.Bleeding was suspected as the patient decompensated.The patient was then brought back to the or, placed on bypass again, and the procedure was repeated/escalated to total replacement of the valve in question.Bleeding was confirmed, at that time, and the necessity to redo the procedure.The physician decided not to place patient on ecls again, or attempt another iab.The patient was scheduled to be transferred to another facility, but the tertiary care facility would not accept the patient since they were not on ecls support.Patient was then transferred to the icu where they expired shortly after.This report is for the 1st iab used in this event.A separate report will be submitted for the 2nd iab.An additional report for the cardiosave iabp used in this event has been submitted under mfg report number 2249723-2023-03813.
|