• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT3 III; TOTAL TRIIODOTHYRONINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT3 III; TOTAL TRIIODOTHYRONINE Back to Search Results
Catalog Number 09005803190
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the e601 analyzer is (b)(6).The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with elecsys ft3 iii on a cobas 6000 e601 module.The issue occurred when starting to use a new reagent kit.No questionable results were reported outside of the laboratory.The sample initially resulted in an ft3 value of 32.55 pg/ml.The sample did not match the patient's clinical diagnosis, so the sample was repeated.The repeat result was 2.67 pg/ml.
 
Additional Manufacturer Narrative
The investigation could not identify a product problem.The cause of the event could not be determined.The issue is likely related to foam present on the reagent after replacement of the reagent kits.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT3 III
Common Device Name
TOTAL TRIIODOTHYRONINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17694934
Report Number1823260-2023-02897
Device Sequence Number12323906
Product Code CDP
UDI-Device Identifier07613336171585
UDI-Public07613336171585
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number09005803190
Device Lot Number660556
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/18/2023
Supplement Date Received by Manufacturer09/11/2023
Initial Report FDA Received Date09/06/2023
Supplement Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-