• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD EDS DRAINAGE SYS NO V CATH; CSF DRAIN SYS/CRANIAL ACCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD EDS DRAINAGE SYS NO V CATH; CSF DRAIN SYS/CRANIAL ACCESS Back to Search Results
Catalog Number 821731C
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes High Blood Pressure/ Hypertension (1908); Nausea (1970); Tachycardia (2095)
Date of Event 07/04/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A facility reported an accidental injection of radio contrast medium into the sampling/injection site of the ventricular shunt device in place of the vascular access.The contrast product was therefore injected into the cerebral ventricles.Patient presented with nausea, hypertension, tachycardia.The patient is an adult of approximately 60 years-old.The following days : general tonico-clonic crisis and hydrocephalus meningitis at pseudomonas & candida, septic vasculitis, ischemic stroke + appearance of stenosis.Long term: cognitive problems, left hypoesthesia, total dependence, continence and swallowing problems.Actions taken in the healthcare establishment to manage the patient: immediate management by the intensive care unit team.Patient has incapacitating sequela.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.Other "health effect-clinical code" 2223, 3272, 2389, 4418, 2263, 2551, 2352, 1815.
 
Additional Manufacturer Narrative
The eds drainage system was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue is due to user¿s error, failure to adhere to ifu warning label regarding using the needle free sampling/injection site.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.A medical assessment was requested to our medical safety department with the following conclusion: use error.Failure to adhere to ifu label regarding using the needle-free sampling/injection site."warning" the eds3 system provides direct access to the ventricular drainage catheter via the patient line stopcock and the needle-free injection site.Care must be taken when administering medication through these locations to minimize the chance of intravascular medications being injected into the cerebrospinal fluid.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDS DRAINAGE SYS NO V CATH
Common Device Name
CSF DRAIN SYS/CRANIAL ACCESS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17694985
Report Number3013886523-2023-00304
Device Sequence Number6306434
Product Code JXG
UDI-Device Identifier10381780518327
UDI-Public10381780518327
Combination Product (Y/N)N
PMA/510(K) Number
K061568
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number821731C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2023
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date09/06/2023
Supplement Report FDA Received Date10/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-