• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC PROPHECY INFINITY PATIENT SPECIFIC GUIDES; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC PROPHECY INFINITY PATIENT SPECIFIC GUIDES; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number PROPINF
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Insufficient Information (4580)
Date of Event 08/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed. a review of the device history is not possible because the lot number was not communicated. should additional information become available, it will be provided on a supplemental report.H3 other text : device disposition unknown.
 
Event or Problem Description
It was reported that the patient prophecy guides were incompatible with patient's anatomy after being built.No additional information is available at this time.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.Upon further investigation by the prophecy team, the following was noted: the ct scan was processed within normal, acceptable ranges.No errors were identified during the segmentation procedure.A loose body osteophyte was found on the distal tibia.This body was segmented according to the procedure, and it was recommended to be removed prior guide placement.The alignment guides were designed within normal, acceptable ranges.No abnormalities in the internal process were found.The engineering drawings and quality documents were processed within normal, acceptable ranges.No abnormalities in the internal process were found.It was not possible to identify what caused the tibia prophecy block inaccuracy mentioned in the complaint¿s email.A potential root cause may be a discrepancy in user expectations vs guide fit and feel during surgery or user error during surgery.Another potential root cause could be that the loose osteophyte at the tibia was not properly removed, which might have had an impact on the guide reference.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that the patient prophecy guides were incompatible with patient's anatomy after being built.No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPHECY INFINITY PATIENT SPECIFIC GUIDES
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17695102
Report Number3010667733-2023-00512
Device Sequence Number2608935
Product Code HSN
UDI-Device Identifier00889797057134
UDI-Public00889797057134
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2024
Device Catalogue NumberPROPINF
Device Lot Number1765589
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/11/2023
Supplement Date Received by Manufacturer11/07/2023
Initial Report FDA Received Date09/06/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-