MAUDE Adverse Event Report: ST. JUDE MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES 63 CM LE; TRANSSEPTAL CATHETER INTRODUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES 63 CM LE; TRANSSEPTAL CATHETER INTRODUCER Back to Search Results
Model Number 406840
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This report is to advise of skiving found in the introducer during analysis.
 
Additional Manufacturer Narrative
One 8f fast-cath introducer sheath and dilator were received for evaluation.Resistance was noted near the distal end of the dilator when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise, and skiving was noted at the distal end of the dilator.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Artmt600108998 ver.B, fast-cath transseptal guiding introducer instructions for use (ifu) states, ¿during insertion, always use the stylet to facilitate needle passage through the dilator/sheath assembly.Failure to use the stylet may inhibit advancement of the needle and may result in inadvertent needle puncture of the dilator/sheath assembly or skiving of material from the inner surface of the dilator.¿ the cause of the needle insertion difficulty is consistent with skiving.The cause of the skiving is consistent with not following the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES 63 CM LE
Common Device Name
TRANSSEPTAL CATHETER INTRODUCER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17695108
Report Number3005334138-2023-00389
Device Sequence Number2645725
Product Code DYB
UDI-Device Identifier05414734203340
UDI-Public05414734203340
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model Number406840
Device Lot Number8883677
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2023
Initial Report FDA Received Date09/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-