| Model Number |
F0304 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Erythema (1840); Loss of Range of Motion (2032); Deformity/ Disfigurement (2360); Depression (2361); Weight Changes (2607); Breast Discomfort/Pain (4504); Swelling/ Edema (4577)
|
| Date of Event |
05/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Hologic is submitting this report in accordance with 21.Cfr part 803.The submission is based upon the currently available information.The submission of this report does not represent a confirmation of device malfunction involving the hologic products in the event described.Lot and serial number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a lumpectomy on (b)(6) 2018 in which a biozorb was implanted.The patient had a prior diagnosis of right breast invasive lobular carcinoma.Patient reported that the device ¨stuck out¨of her chest and caused constant pain and discomfort.Patient reported that she had difficulty sleeping due to pain and could not hug other people.Patient reported that a physician suggested that the device would dissolve in two years, the patient awaited additionally three years for the device to reabsorb and eventually had a surgery to remove the biozorb in (b)(6)2022.Patient reported relief after the surgery.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
After additional medical review it was determined that a 58-year-old caucasian patient with a body mass index of 33 and 209 pounds, received a biozorb implant (f0304 sutured in place with vicryl3.0 in the upper inner quadrant right breast with breast parenchyma reconstruction over the implant) on (b)(6) 2018.The patient received the surgery due to a previous diagnosis of invasive ductal carcinoma grade 3, pt1c pn0; er/pr +; her-2 negative.Patient underwent a lumpectomy and biozorb placement and reconstruction of deficit.Specimen taken on day of surgery had positive margins and patient required a re-excision on (b)(6) 2018 and the biozorb marker was left in place.The pathology results after the re-excision revealed no residual tumor.When seen in office 3 weeks after the re-excision patient complained of pain and swelling and erythema at the site was noted.Needle aspiration of seroma was reported and 5 days after the drainage erythema was noted at the site and patient was started on antibiotics (bactrim and clindamycin).Erythema reported as resolved prior to radiation.2 months after the re-excision surgery patient received physical therapy for discomfort and decreased range of motion and tenderness at the scar.Patient underwent adjuvant radiotherapy from (b)(6) 2018 and 4 months after radiotherapy patient reported chest wall tenderness and that the biozorb was palpable.Patient reported depression and weight gain after these events and reported persistent discomfort at lumpectomy site and the surgeon reported tethering of right breast from biozorb to nipple.Patient was unhappy with the breast deformity and was considering explant of the device and desired a breast reduction at 2 years post surgery.At 3 years biozorb was still palpable and patient underwent a fine needle biopsy due to a nodule on her right breast which revealed benign fatty necrosis.Patient requested the removal and the biozorb device was explanted on (b)(6) 2024 and received a fat graft to fill the defect.The pathology analyzed after the explantation revealed the following: right areolar scar: hypertrophic dense scar tissue; right breast upper inner quadrant: fibroadipose tissue with dense fibrous tissue, fat necrosis, dystrophic calcification, mixed inflammation, foamy histiocytes and foreign body giant cells.No note of biozorb in pathology report from explant.Patient has a previous history of diabetes mellitus, hypertension, thyroid problems, hyperlipidemia, hysterectomy and right oophorectomy for endometriosis, and anxiety.Maternal aunt with breast cancer at age 80 and sister at age 66.Due to her cancer diagnosis patient received hormone therapy: anastrazole to be taken for 5 years and radiation to ipsilateral breast from (b)(6) 2018 to.Patient did not receive chemotherapy and no cancer recurrence was reported.Lot number was provided and a device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed: it was reported that a patient underwent a lumpectomy on (b)(6) 2018, in which a biozorb was implanted.The patient had a prior diagnosis of right breast invasive lobular carcinoma.Patient reported that the device ¨stuck out¨of her chest and caused constant pain and discomfort.Patient reported that she had difficulty sleeping due to pain and could not hug other people.Patient reported that a physician suggested that the device would dissolve in two years, the patient awaited additionally three years for the device to reabsorb and eventually had a surgery to remove the biozorb in (b)(6) 2022.Patient reported relief after the surgery.Emotional distress.No initial evaluation could be performed because there was no returned sample for this complaint.The sample was not returned for this complaint and there was limited patient-related information provided for this event; therefore, a thorough investigation was not able to be performed.Investigation into several of the harms reported in this complaint was performed from a clinical literature review, as well as an investigation into the design and risk of the biozorb product.The harms identified during this complaint are: major pain (pain/discomfort/device palpability/sleep disfunction/failure to resorb), minor inflammation (inflammation/swelling/lymphedema/limited mobility), moderate tissue damage (tissue damage and/or skin erosion), major additional intervention (additional surgery/device explantation/calcifications).A review of clinical literature performed on lumpectomies and their associated complications was performed.A summary of this review concluded that, while lumpectomy is a well-established option for breast cancer treatment, it inherently carries risks of complications, most notably infections, seroma formation, chronic pain, and wound healing issues.The literature demonstrates variation in complication rates based on surgical technique, additional therapies (such as intraoperative radiation therapy (iort) or radiation), and individual patient factors, with infection rates ranging from 2% to 10% and seroma rates from 11% to 65%.Chronic pain, particularly neuropathic pain, affects nearly one-third of patients.(kwee et al.(2024)), migration can be influenced by several patient related factors which can influence the likelihood of migration after lumpectomy, such as breast density, biopsy location, and biopsy approach.Per a study by ashali jain et al.(2017), marker migration is a relatively common occurrence following stereotactic biopsies, happening in 13% of cases.Swelling is one of the expected events that could occur after surgery and/or radiation therapy regardless of the use of a biozorb device.Due to this, confirmation of the swelling during this event as a normal course of surgery vs being caused by the biozorb is not able to be done.Device explantation is based on the patient¿s physician¿s evaluation of their physical and health state and recommendation; therefore, an investigation into this intervention is not able to be performed for this complaint.Though complications are relatively common, most are manageable with appropriate postoperative care, making lumpectomy a viable option for many breast cancer patients when carefully monitored and followed up.An investigation was performed by subject matter experts (sme¿s) into several of the harms identified as being experienced by the patient related to design, risk management, and clinical factors.The results of these investigations and potential causes can be seen below; however, it is important to note that these results do not include medical input from a qualified medical professional: pain: there is a lack of migration requirements and testing to evaluate if the device can cause tissue damage due to migration.Verification of mechanical strength and simulated use is not adequate due to several opportunities related to sampling size and test methods.The crystalline proportion in the device is unknown.Crystallinity may be causing irritation and foreign body reaction considering the long period of absorption of the device.There is a lack of usability testing evaluating the user needs, so there is no assurance that the surgical methods being used for the users are adequate.There is a lack of adequate user needs to capture appropriate design inputs, and therefore there are opportunities in design outputs and v&v activities.Migration: there is a lack of migration requirements and testing to evaluate if the device can cause tissue damage due to migration.The crystalline proportion in the device is unknown.High crystallinity may be causing migration/erosion considering the long period of absorption of the device.Verification of mechanical strength and simulated use is not adequate due to several opportunities related to sampling size and test methods.There is a lack of usability testing evaluating the user needs, so there is no assurance that the surgical methods being used for the users are adequate.There is a lack of adequate user needs to capture appropriate design inputs.Based on the clinical literature review, the harms reported to be experienced by the patient in the complaint are common symptoms from surgery and lumpectomies; therefore, a confirmation of the biozorb causing the harms was not able to be done.However, based on the review of the top harms in regards to the biozorb design and risk documentation, the adverse events reported can be tracked back to a potential shared cause: lack of adequate user needs for the product.User needs define the framework for product development and lay the ground for the definition of design inputs, design outputs and later verification and validation activities.Other potential causes are related to inadequate test methods and/or the lack of evidence to support some specific product requirements.Conclusion: when comparing the safety outcomes of the biozorb marker with general lumpectomy complications, it becomes evident that some of the complications associated with biozorb may stem from the lumpectomy procedure itself.Both procedures share overlapping risks, although the biozorb device introduces additional considerations.However, it has been confirmed that the lack of defined user needs and inadequate test methods and testing evidence are potential root causes for the harms reported in this complaint.This cause will be further investigated and addressed in a capa.The risks associated with this complaint event are further described and evaluated against the product and its hazard analysis in the health risk assessment biozorb complaint analysis.Updates to the biozorb hazard analysis risk file will be monitored.Future events will be monitored and trended.Complaint confirmed: no.Device history record (dhr) review: a device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
|
| |
|
Event or Problem Description
|
|
Hologic has received correspondence arising from litigation.The additional medical information provided, revealed the explant date of the device as (b)(6) 2022.No other relevant information was provided.This report is being submitted pursuant to 21 cfr part 803 based on information made available to hologic through litigation.Consistent with 21 cfr 803.16, the submission of this report is not intended, and shall not be construed, as an admission, acknowledgment, or confirmation by hologic that the device caused or contributed to the reportable event.
|
| |
|
Additional Manufacturer Narrative
|
|
Hologic has received correspondence arising from litigation.The additional medical information provided, revealed the explant date of the device as (b)(6) 2022.No other relevant information was provided.This report is being submitted pursuant to 21 cfr part 803 based on information made available to hologic through litigation.Consistent with 21 cfr 803.16, the submission of this report is not intended, and shall not be construed, as an admission, acknowledgment, or confirmation by hologic that the device caused or contributed to the reportable event.
|
| |
|
Search Alerts/Recalls
|
|
|