It was reported that a patient underwent a right partial mastectomy on (b)(6) 2020 during which a biozorb was implanted.Patient had a prior diagnosis of right breast invasive lobular carcinoma.Patient reported suffered from pain, a hard lump, deformity, scarring, sensitivity, itching, swelling and reddening of her skin, fibrosis, lymphedema, seroma, and edema.Patient reported is having issues sleeping and causing discomfort.No initial evaluation could be performed because there was no returned sample for this complaint.The sample was not returned for this complaint and there was limited patient-related information provided for this event; therefore, a thorough investigation was not able to be performed.Investigation into several of the harms reported in this complaint was performed from a clinical literature review, as well as an investigation into the design and risk of the biozorb product.The harms identified during this complaint are: pain, infection, seroma, redness/erythema, itching, deformity, swelling, sleep dysfunction, lymphedema a review of clinical literature performed on lumpectomies and their associated complications was performed.A summary of this review concluded that, while lumpectomy is a well-established option for breast cancer treatment, it inherently carries risks of complications, most notably infections, seroma formation, chronic pain, and wound healing issues.The literature demonstrates variation in complication rates based on surgical technique, additional therapies (such as intraoperative radiation therapy (iort) or radiation), and individual patient factors, with infection rates ranging from 2% to 10% and seroma rates from 11% to 65%.Pain biozorb (0.9% to 1.9%): pain or discomfort was reported in 0.9% to 1.9% of patients with the biozorb marker.In one instance, discomfort occurred due to the device being sutured to the serratus muscle but did not result in removal.In another study, discomfort in 1.1% of patients led to device removal.However, given that pain is a well-known complication following lumpectomy, particularly in cases with nerve damage or radiation therapy, some of this discomfort could be attributed to the initial surgical intervention rather than the presence of the marker.Lumpectomy (25% to 60%): chronic pain is a frequent long-term complication of lumpectomy, affecting 25% to 60% of breast cancer surgery survivors.Kwee et al.Reported a pooled prevalence of 31% for neuropathic pain following bcs.The presence of post-surgical pain, particularly neuropathic pain caused by nerve damage, could be exacerbated by or misattributed to the biozorb marker in some cases.Infection biozorb (0% to 5.6%): infection rates associated with biozorb ranged from 0% to 5.6%, with 1.9% of patients across studies experiencing infections.In some cases, infections were severe enough to necessitate device removal, with 5 out of 17 infections leading to removal in a registry study.Given that infection is a known complication of lumpectomy, it is likely that some infections reported in biozorb studies are related to the surgical trauma from the lumpectomy procedure, especially in cases involving re-excisions or extensive tissue manipulation.Lumpectomy (2% to 10%): infection is a recognized complication of lumpectomy, with rates ranging from 2% to 10%.Fernando et al.Reported an infection rate of 2.6% for traditional lumpectomy, while boniface et al.Found infection rates of 2.1% for breast-conserving surgery (bcs), rising to 4.6% for those undergoing mastectomy.Seroma biozorb (0.7% to 3.3%): seroma formation rates with biozorb ranged from 0.7% to 3.3%.Kaufman et al.(2021)4 observed that 82.8% of seromas occurred within the first three months after surgery, suggesting a link to the initial surgical procedure rather than the device itself.The overall seroma rate was 2.8%.Given that seromas frequently develop in response to the space left after lumpectomy, the seroma formation observed in biozorb studies may be more related to the lumpectomy rather than the device.Lumpectomy (up to 65%): seroma is a common complication following lumpectomy, with reported rates as high as 65% in some studies, particularly when combined with intraoperative radiotherapy (iort) or axillary procedures.Sfarad et al.Reported a 65% seroma rate following iort, suggesting that the surgical removal of tissue, not the marker itself, is the primary driver of this complication.Erythema / redness (hypersensitivity/allergic reaction/ itching sensation) claudia et al (admoun c, mayrovitz h.Choosing mastectomy vs.Lumpectomy-with-radiation: experiences of breast cancer survivors.Cureus.Oct 2021;13(10): e18433.Doi:10.7759/cureus.18433) explored the experiences of 1,606 breast cancer survivors who underwent either mastectomy or lumpectomy with radiation, focusing on safety, complications, and patient satisfaction.Of the respondents, 978 had mastectomies and 628 had lumpectomies.The study found that lumpectomy patients experienced higher rates of radiation-related side effects, with 98% reporting issues such as skin irritation, thickening, and chest wall tenderness.Sfarad et al.(sfarad hk, allweis tm.Postoperative complications following lumpectomy with intraoperative x-ray radiation therapy: a retrospective comparative study.Clin breast cancer.Apr 2024;24(3):237-242.Doi: 10.1016/j.Clbc.2023.12.005) compared complication rates in patients undergoing lumpectomy with intraoperative radiation therapy (iort), external beam radiation therapy (ebrt), or lumpectomy alone for early breast cancer.Among the patients, seroma occurred in 65% after iort, 34% after ebrt, and 11% after lumpectomy alone (p =.000).Surgical site infections were diagnosed in 23% after iort, 15% after ebrt, and 6% after lumpectomy alone (p =.013).Postoperative erythema was reported in 34% after iort, 16% after ebrt, and 6% after lumpectomy alone (p =.000).Minor complications like scar formation, edema, and chronic tenderness occurred in 55% after iort, 47% after ebrt, and 17% after lumpectomy alone (p =.000).The study concludes that iort is associated with a higher rate of complications compared to ebrt and lumpectomy alone, although most complications were minor and transient.The authors suggest that the increased complication rates may be partially due to overreporting with the introduction of a new technology.Educating physicians and patients about potential complications and their course is recommended to help manage postoperative outcomes.Deformity (physical asymmetry / disfigurement) rahimi 2022 (rahimi a, simmons a, kim dn, et al.Preliminary results of multi-institutional phase 1 dose escalation trial using single-fraction stereotactic partial breast irradiation for early-stage breast cancer.Int j radiat oncol biol phys.Mar 1, 2022;112(3):663-670.Doi: 10.1016/j.Ijrobp.2021.10.010) concluded that cosmetic outcomes were preserved, with no significant cosmetic detriment across any dose cohort, suggesting that single-fraction s-pbi does not negatively affect patients' physical appearance post-treatment.Both patient- and physician-reported cosmetic scores were consistently favorable over 12 and 24 months of follow-up.Swelling / lymphedema (inflammation / limited mobility) biozorb (0.7%): tissue damage, including wound dehiscence and erosion, was reported in 0.7% of patients with the biozorb marker.In a more severe case, the biozorb device eroded into the nipple-areola complex, necessitating complete removal of the nipple-areola complex due to chronic inflammation.However, given that wound healing issues, tissue necrosis, and other damage are known complications of lumpectomy itself, some of these outcomes could be directly related to the surgical trauma and healing process rather than the device.Lumpectomy (13.1% of patients, delayed healing in 6.5%, and skin-flap necrosis in 1.1%, indicating that tissue damage): wound healing issues and tissue necrosis are common after lumpectomy, particularly in cases where large amounts of tissue are removed or additional procedures such as radiation are performed.Knowles et al.Reported wound infection in 13.1% of patients, delayed healing in 6.5%, and skin-flap necrosis in 1.1%, indicating that tissue damage can be a lumpectomy-related issue.Regarding lumpectomy¿s, berger et al.(berger l, grimm a, sutterlin m, et al.Major complications after intraoperative radiotherapy with low-energy x-rays in early breast cancer.Strahlenther onkol.Apr 2024;200(4):276-286.Doi: 10.1007/s00066-023-02128-z) retrospectively analyzed the occurrence of severe local complications in 10 out of 408 women who underwent intraoperative radiotherapy (iort) with low-energy x-rays during breast-conserving surgery (bcs) for early breast cancer between 2002 and 2017.The complications, which required surgical intervention, included redness (80%), seroma (60%), wound infection (60%), suture dehiscence (60%), and induration (40%).Hematoma and necrosis were less common (10% each).These symptoms emerged from immediately post-operation up to 15 years later, with a median onset at 3.1 months.While most complications were managed with smaller surgical interventions, such as excision of necrotic areas or secondary sutures, more severe cases required complex flap surgery or mastectomy.The study concluded that although iort is generally safe, it carries a risk of severe complications.Preoperative counseling and vigilant follow-up are recommended to manage these risks effectively.An investigation was performed by subject matter experts (sme¿s) into several of the harms identified as being experienced by the patient related to design, risk management, and clinical factors.The results of these investigations and potential causes can be seen below; however, it is important to note that these results do not include medical input from a qualified medical professional: pain: there is a lack of migration requirements and testing to evaluate if the device can cause tissue damage due to migration.Verification of mechanical strength and simulated use is not adequate due to several opportunities related to sampling size and test methods.The crystalline proportion in the device is unknown.Crystallinity may be causing irritation and foreign body reaction considering the long period of absorption of the device.There is a lack of usability testing evaluating the user needs, so there is no assurance that the surgical methods being used for the users are adequate.There is a lack of adequate user needs to capture appropriate design inputs, and therefore there are opportunities in design outputs and v&v activities.Migration: there is a lack of migration requirements and testing to evaluate if the device can cause tissue damage due to migration.The crystalline proportion in the device is unknown.High crystallinity may be causing migration/erosion considering the long period of absorption of the device.Verification of mechanical strength and simulated use is not adequate due to several opportunities related to sampling size and test methods.There is a lack of usability testing evaluating the user needs, so there is no assurance that the surgical methods being used for the users are adequate.There is a lack of adequate user needs to capture appropriate design inputs.Based on the clinical literature review, the harms reported to be experienced by the patient in the complaint are common symptoms from surgery and lumpectomies; therefore, a confirmation of the biozorb causing the harms was not able to be done.However, based on the review of the top harms in regard to the biozorb design and risk documentation, the adverse events reported can be tracked back to a potential shared cause: lack of adequate user needs for the product.User needs define the framework for product development and lay the ground for the definition of design inputs, design outputs and later verification and validation activities.Other potential causes are related to inadequate test methods and/or the lack of evidence to support some specific product requirements.Conclusion: when comparing the safety outcomes of the biozorb marker with general lumpectomy complications, it becomes evident that some of the complications associated with biozorb may stem from the lumpectomy procedure itself.Both procedures share overlapping risks, although the biozorb device introduces additional considerations.However, it has been confirmed that the lack of defined user needs and inadequate test methods and testing evidence are potential root causes for the harms reported in this complaint.This cause will be further investigated and addressed in a capa.The risks associated with this complaint event are further described and evaluated against the product and its hazard analysis in the health risk assessment biozorb complaint analysis.Updates to the biozorb hazard analysis risk file will be monitored.Future events will be monitored and trended.Complaint confirmed: no.Device history record (dhr) review: a dhr review could not be performed because no lot information was provided.
|