| Catalog Number |
68-180 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b.5., d6b., h.6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: not observed.¿ creases/folding of implant: observed, flat creases.Additional observations: ¿ white particles in the device inner surface are observed.No further actions are required since no issue in the manufacturing of the device is observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Healthcare professional later reported "deflate-crease fold failure".Device has been explanted and replaced.
|
| |
|