BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 368652
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 08/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D.4.Medical device expiration date: unknown.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.H.4.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that during use with bd vacutainer® safety-lok¿ blood collection set with pre-attached holder air bubbles formed and blew patient's vein.There was no additional impact reported.The following information was provided by the initial reporter: customer reports cat # 368652 had started drawing air with the blood, going super, and blew the patient's vein.
 
Additional Manufacturer Narrative
H3.Investigation summary: bd had not received samples or photos for investigation.Therefore, 30 retention samples from bd inventory were evaluated by functional draw testing and no issues were observed relating to air bubbles as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode air bubbles.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends h3 other text : see h10.
 
Event or Problem Description
It was reported that during use with bd vacutainer® safety-lok¿ blood collection set with pre-attached holder air bubbles formed and blew patient's vein.There was no additional impact reported.The following information was provided by the initial reporter: customer reports cat # 368652 had started drawing air with the blood, going super, and blew the patient's vein.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17696059
Report Number1024879-2023-00613
Device Sequence Number12323917
Product Code JKA
UDI-Device Identifier50382903686525
UDI-Public50382903686525
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K980414
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number368652
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2023
Supplement Date Received by Manufacturer10/13/2023
Initial Report FDA Received Date09/06/2023
Supplement Report FDA Received Date10/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-