The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the electrical pin connector fluid damage, the lg connector fluid damage, the lg connector corroded, the remote control buttons cut, the distal body worn out, the insertion flexible tube lump/wave, the suction channel kink, the angulation down angulation zero degrees, the up pulley wire snapped/cut, and the u/d and the r/l knobs white marking faded/missing; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|