Catalog Number AK-04510-S
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 08/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that: the guidewire gets stuck in the introducer.They then retract the guidewire and use the harder side of the tip or put in another line.Additional information: the guide wire was found to be kinked upon removal.The insertion attempts were performed under ultrasound.There was no scar tissue present and there were no issues advancing the wire in the vessel.Arterial blood flow was present prior to advancing the guidewire.The consequence to the patient of the incident was arterial hematoma that requires manual pressure.An inability to gain accurate blood pressure readings until another line could be placed successfully.The patient was reportedly in severe cardiogenic shock.Additional information has been requested from the account and will be added to the follow up report.The patient's current condition is reported as critical but stable , but not reported as a consequence of the incident.The issue was found in 8 different incidents with 8 different patients.Additional complaints are associated: (b)(4).
 
Additional Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "warning: to reduce the risk of guidewire damage, do not retract guidewire against edge of needle while in vessel." a device history record review was performed based on a potential lot from sales history, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable root cause could not be determined from the available information.No further action was required at the time of the report.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that: the guidewire gets stuck in the introducer.They then retract the guidewire and use the harder side of the tip or put in another line.Additional information: the guide wire was found to be kinked upon removal.The insertion attempts were performed under ultrasound.There was no scar tissue present and there were no issues advancing the wire in the vessel.Arterial blood flow was present prior to advancing the guidewire.The consequence to the patient of the incident was arterial hematoma that requires manual pressure.An inability to gain accurate blood pressure readings until another line could be placed successfully.The patient was reportedly in severe cardiogenic shock.Additional information has been requested from the account and will be added to the follow up report.The patient's current condition is reported as critical but stable , but not reported as a consequence of the incident.The issue was found in 8 different incidents with 8 different patients.Additional complaints are associated: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW ARTERIAL LINE KIT
Common Device Name
WIRE GUIDE CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key17698333
Report Number9680794-2023-00695
Device Sequence Number15331141
Product Code DQX
UDI-Device Identifier20801902121688
UDI-Public20801902121688
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K810675
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberAK-04510-S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2023
Supplement Date Received by Manufacturer09/29/2023
Initial Report FDA Received Date09/07/2023
Supplement Report FDA Received Date10/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-