OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TCRIS; ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES Back to Search Results
Model Number WA47706S
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/09/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported to olympus the hf-resection electrode "plasmaloop" was missing after being used in a patient during a hysteroscopy with fibroid resection.The loop was there when inserting device and when removing, the loop was missing.It was stated that the loop broke inside the patient and was not retrievable, and it potentially was retained by the patient.An x-ray after the procedure and a thorough examination of the uterus was done, but they were unable to visualize or locate the missing portion.It was stated that the procedure was unable to be completed.Additional follow up was requested.
 
Additional Manufacturer Narrative
The device was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and updates to b5, d4, d9, h3, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.The exact root cause of the reported phenomenon could not be conclusively determined.It has been concluded that the possible cause of the event was likely due wear and tear.Despite the noticeable interruption of the loop wire, the user continued to resect tissue until the wire had burned down completely.To mitigate the risk of device malfunction and patient/user injury, the instruction manual contains the following description related to this reported phenomenon: caution 1: "caution risk of injury to the patient contact between the hf-resection electrode and metal parts, such as other endoscopic equipment, implants or stents can result in sparkover between the hf-resection electrode and the metal part.Always keep a distance of at least 10 mm between the hf-resection electrode and metal parts." caution 2: "caution risk of injury to the patient use extreme caution when using electrosurgery in close proximity to or in direct contact with any metal objects.Do not activate the hf-resection electrode while any portion of the hf-resection electrode tip is in contact with another metal object.This can cause uncontrolled heating of the hf-resection electrode and may result in damage and breakage to the hf-resection electrode tip.If excessive heating or physical forces cause damage to the hf-resection electrode tip, broken device fragments may remain in the body, possibly requiring additional surgery for removal." olympus will continue to monitor field performance for this device.
 
Event or Problem Description
Additional information stated that the patient was discharged as planned, and there was no extension of their hospital stay.It was noted that the device was inspected prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TCRIS
Common Device Name
ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17699301
Report Number9610773-2023-02484
Device Sequence Number17991971
Product Code FAS
UDI-Device Identifier14042761085530
UDI-Public14042761085530
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171965
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2025
Device Model NumberWA47706S
Device Lot Number1100410974
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2023
Supplement Date Received by Manufacturer08/30/2024
Initial Report FDA Received Date09/07/2023
Supplement Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-