| Model Number |
WA47706S |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported to olympus the hf-resection electrode "plasmaloop" was missing after being used in a patient during a hysteroscopy with fibroid resection.The loop was there when inserting device and when removing, the loop was missing.It was stated that the loop broke inside the patient and was not retrievable, and it potentially was retained by the patient.An x-ray after the procedure and a thorough examination of the uterus was done, but they were unable to visualize or locate the missing portion.It was stated that the procedure was unable to be completed.Additional follow up was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and updates to b5, d4, d9, h3, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.The exact root cause of the reported phenomenon could not be conclusively determined.It has been concluded that the possible cause of the event was likely due wear and tear.Despite the noticeable interruption of the loop wire, the user continued to resect tissue until the wire had burned down completely.To mitigate the risk of device malfunction and patient/user injury, the instruction manual contains the following description related to this reported phenomenon: caution 1: "caution risk of injury to the patient contact between the hf-resection electrode and metal parts, such as other endoscopic equipment, implants or stents can result in sparkover between the hf-resection electrode and the metal part.Always keep a distance of at least 10 mm between the hf-resection electrode and metal parts." caution 2: "caution risk of injury to the patient use extreme caution when using electrosurgery in close proximity to or in direct contact with any metal objects.Do not activate the hf-resection electrode while any portion of the hf-resection electrode tip is in contact with another metal object.This can cause uncontrolled heating of the hf-resection electrode and may result in damage and breakage to the hf-resection electrode tip.If excessive heating or physical forces cause damage to the hf-resection electrode tip, broken device fragments may remain in the body, possibly requiring additional surgery for removal." olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
Additional information stated that the patient was discharged as planned, and there was no extension of their hospital stay.It was noted that the device was inspected prior to use.
|
| |
|