• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FLEXIBLE REAMER 6.0 MM DIAMETER; ORTHOPAEDIC PROSTHESIS IMPLANTATION POSITIONING INSTRUMENT, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FLEXIBLE REAMER 6.0 MM DIAMETER; ORTHOPAEDIC PROSTHESIS IMPLANTATION POSITIONING INSTRUMENT, REUSABLE Back to Search Results
Catalog Number 00222800600
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-02442.D10: item# 00222800500; lot# 62537293.G2: foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product cannot be returned due to hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during a procedure, the 5mm reamer disconnected from the proximal connector and the 6mm reamer shattered inside the tibia.The fractured pieces of the 6mm reamer were unable to be removed and have been retained by the patient.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: proposed component code - mechanical (g04) - drill.Visual examination of the provided pictures identified the 6.0mm reamer to be fractured into two pieces and the 5.0mm reamer to be disassembled into two pieces.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: there are osseous findings consistent with osteogenesis imperfecta.Operative changes are present with an im mail within the tibia and there are numerous small metallic fragments as well as a larger linear metallic fragment in the proximal tibia consistent with the reported double reamer failure.No displaced osseous fracture is noted.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE REAMER 6.0 MM DIAMETER
Common Device Name
ORTHOPAEDIC PROSTHESIS IMPLANTATION POSITIONING INSTRUMENT, REUSABLE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17699787
Report Number0001822565-2023-02441
Device Sequence Number15315031
Product Code HWE
UDI-Device Identifier00889024038738
UDI-Public(01)00889024038738(10)63324010
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00222800600
Device Lot Number63324010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2023
Supplement Date Received by Manufacturer01/05/2024
Initial Report FDA Received Date09/07/2023
Supplement Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H10.
Outcome Attributed to Adverse Event Other;
Patient Age16 YR
Patient SexMale
-
-