• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH CARBON DIOXIDE; ENZYMATIC, CARBON-DIOXIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH CARBON DIOXIDE; ENZYMATIC, CARBON-DIOXIDE Back to Search Results
Catalog Number 03L80-22
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A final report will be submitted when the evaluation is complete.Completed information for section a1 patient identification: (b)(6).All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer observed falsely elevated carbon dioxide results generated on the architect c4000 analyzer for multiple patients.The following data was provided (normal range 22-29 meq/l): (b)(6).No impact to patient management was reported.
 
Additional Manufacturer Narrative
Updated information in sections: d4 - expiration date updated from blank to 3/31/2024 , g1, h4 - device mfg date updated from blank to 3/3/2023 the complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue.Ticket search by lot indicates that the complaint lot performs as expected.Ticket trending review did not identify any trends.Device history review did not identify any non-conformances or deviations with the complaint lot.Labeling was reviewed and adequately addresses the current issue.Based on the investigation, no systemic issue or deficiency of the architect carbon dioxide reagent, lot 63532uq02, was identified.
 
Event or Problem Description
The customer observed falsely elevated carbon dioxide results generated on the architect c4000 analyzer for multiple patients.The following data was provided (normal range 22-29 meq/l): sample id (b)(6) initial result = 30.24 mmol/l, repeat = 25.23 mmol/l sample id (b)(6) initial result = 33.15 mmol/l, repeat = 24 mmol/l sample id (b)(6) initial result = 29.87 mmol/l, repeat = 22.54 mmol/l sample id (b)(6) initial result = 33.45 mmol/l, repeat = 24 mmol/l sample id (b)(6) initial result = 31.28 mmol/l, repeat = 19.1 mmol/l no impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBON DIOXIDE
Common Device Name
ENZYMATIC, CARBON-DIOXIDE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key17701939
Report Number3002809144-2023-00384
Device Sequence Number6718937
Product Code KHS
UDI-Device Identifier00380740161521
UDI-Public00380740161521
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K060295
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Catalogue Number03L80-22
Device Lot Number63532UQ02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/21/2023
Supplement Date Received by Manufacturer10/06/2023
Initial Report FDA Received Date09/07/2023
Supplement Report FDA Received Date10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/03/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4000 INTGR, 02P24-40, C401915; ARC C4000 INTGR, 02P24-40, C401915
Patient SexUnknown
-
-