|
SYNTHES GMBH REDUCTION DRIVE UNIT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
511.785 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was not returned for evaluation.Quality engineering evaluated the reduction drive device and the reported condition that the device was not working was confirmed.An assessment was performed and it was determined that the device was not functioning at all when the motor was working.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi - (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from switzerland that during service and evaluation, it was determined that the reduction drive device failed visual inspection due to the component was damage and not moving properly.An assessment was performed and it was found that the cutter/attachment did not show signs of movement of rotation / oscillation / reciprocation when the motor was working due to which the device does not function at all.It was further determined that the device failed pretest for general condition, check for free movement, and check for only forward function.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|