| Model Number |
ZIT-520-5 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pressure bag leaking pressure could delay or interrupt therapy.This is attached to an arterial line and can impact patient therapy.
|
| |
|
Event or Problem Description
|
|
Infusor leaked air within 10 minutes of beginning use.
|
| |
|
Event or Problem Description
|
|
Infusor leaked air within 10 minutes of beginning use.
|
| |
|
Additional Manufacturer Narrative
|
|
The pressure bag leaking pressure could delay or interrupt therapy.This is attached to an arterial line and can impact patient therapy.Complaint confirmed by customer's photo of defect.Inventory inspection performed and no non-conformances were identified.The root cause is likely that not enough glue was used.Performed risk assessment and determined a severity rating of 6/10.Sent resolution to the customer.
|
| |
|