| Catalog Number |
21-4035-24 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Extravasation (1842)
|
| Date of Event |
08/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other text: d4 and g5 are unknown.No product information has been provided for these fields.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
A customer stated that there was an incident with device, which was not discovered prior to patient use and the event result in patient harm in extravasation of chemotherapy.Medical intervention was required where surgical removal and subcutaneous wash out.The outcome of the event is still ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6.Evaluation codes: updated.Device evaluation: one sample was received in used condition.A visual inspection found no damage.During the functional testing, liquid flowed through the product without a problem.As an additional test the connection between the port and the catheter was verified.No discrepancies were found, the test successfully passed.The complaint was unable to be confirmed.A dhr review was completed and no non-conformances were found.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|