Model Number 10848280
Medical Device Problem Code Operating System Becomes Nonfunctional (2996)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
Siemens became aware of a malfunction that occurred while operating the artis q floor system, during an emergency procedure, the system froze and was not operable.The procedure had to be cancelled by the user.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
H3, h6: the investigation was performed based on expert discussions considering complaint description, customer service reports, system history, and system log files.According to the provided information, x-ray was not possible during an emergency procedure.No health patient consequence was reported to this event.Log-file investigation showed that x-ray was possible in fluoro-mode.However, ¿store fluoro¿ and x-ray was not possible in acquisition mode.The cause is found to be that the image disk was full and there was not enough space to creating image file, which resulted in no images being saved.As the imaging system is not an archiving system, deletion of old patient data periodically is required to avoid full memory.The message "memory full, delete patients" message was displayed to the user.The instructions for use provide adequate guidance for proper use in the event of this message.To resolve the problem, old patient data was deleted.No malfunction of the system could be detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS Q FLOOR
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1 or
rittigfeld 1
forchheim 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1
or rittigfeld 1
forchheim 91301
GM   91301
Manufacturer Contact
rebecca tudor
40 liberty blvd., mc 65-1a
malvern, PA 19355
4843234198
MDR Report Key17710365
Report Number3004977335-2023-00109
Device Sequence Number16681633
Product Code OWB
UDI-Device Identifier04056869009971
UDI-Public04056869009971
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K181407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10848280
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2023
Supplement Date Received by Manufacturer09/29/2023
Initial Report FDA Received Date09/08/2023
Supplement Report FDA Received Date09/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-