| Model Number |
SUP/POST AUG PLATE, R RS GLENOID BASEPLATE |
| Medical Device Problem Codes |
Patient Device Interaction Problem (4001); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
07/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.D10: serial number; item number and full description; (b)(6) 320-10-00 - equinoxe reverse tray adapter plate tray +0.(b)(6) 320-42-00 - equinoxe reverse 42mm humeral liner +0.(b)(6) 300-30-06 - equinoxe preserve stem 6mm.(b)(6) 320-06-42 - glenosphere 42mm.
|
| |
|
Event or Problem Description
|
|
As reported by the exactech equinoxe shoulder study, the patient had an initial right tsa on (b)(6) 2020.The patient presented on (b)(6) 2022 with aseptic glenoid loosening and was revised to non-exactech devices on (b)(6) 2022.The case report form indicates this event is definitely related to the device and procedure.This event report was received through clinical data collection activities.Outcome: resolution date: (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported may be the result of the glenoid loosening as reported.However, the loosening cannot be confirmed and any potential contributions from manufacturing, user, or patient-related contributing factors to the event cannot be evaluated as no radiograph, images, or clinical information was provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|