| Catalog Number |
94671 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during final tightening of a small polaris cross connector, the set screw appeared to be stripped; no audible click was heard.The implant was taken off and a new one was used.There was no reported patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed that the device was returned disassembled, however the device was able to be reassembled without incident.All three screws of the assembly were inspected and there was no damage to the threads.Each screw was able to be fully tightened without issue.Root cause: a definitive root cause cannot be determined since the returned device was found to function as expected.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during final tightening of a small polaris cross connector, the set screw appeared to be stripped; no audible click was heard.The implant was taken off and a new one was used.There was no reported patient impact.
|
| |
|