| Model Number |
0998-UC-3023-53 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during preventative maintenance by getinge field service engineer, the cs300 intra-aortic balloon pump (iabp) unit failed battery runtime test.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional info: contact person(name: (b)(6)) it was reported that during preventive maintenance done by a getinge field service engineer (fse) the cs300 intra-aortic balloon pump (iabp) had failed battery run time test.Fse evaluated the unit and stated that battery runtime of 63 minutes.Minimum runtime spec is 135 minutes.Hense fse resolved the issue by replacing d146-00-0039 battery, sealed lead acid,12v.Fse then performed functional testing and device passed testing.Iabp was then returned to customer for clinical use.The defective components were received for further investigation.The failure analysis and testing dept.Received (2) batteries with a reported unit failure of a failed battery runtime test at 63 minutes.The fat performed a visual inspection and found the part to be in good condition.The fat installed the parts into cs300 test fixture and tested the part to factory specifications and the cs300 service manual.The battery runtime test failed at 54 minutes.Fat was able to verify the reported issue.Retaining the part in the failure analysis and testing department per procedure number 0002-07-d008 rev.Ap.The non-conformances with the returned components were confirmed.However, the root cause or the most probable root cause is impossible to be defined.
|
| |
|
Search Alerts/Recalls
|