E1: customer phone = (b)(6).E3: occupation = lead fetal medicine midwife, fetal medicine.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, a 'harrison fetal bladder stent set' was attempted to be used for thoracentesis and pleuro-amniotic shunt placement.During placement, the user attempted to release the pigtail from the trocar twice but the pigtail would not release ((b)(4)/lot 15109821)).The procedure could not be completed, resulting in the need for 3 additional procedures to take place at later, unspecified, dates to drain the excess pleural fluid ((b)(4)/lot 14000403).
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary as reported, a 'harrison fetal bladder stent set' was attempted to be used for thoracentesis and pleuro-amniotic shunt placement.During placement, the user attempted to release the pigtail from the trocar twice but the pigtail would not release (patient id (b)(6)).The procedure could not be completed, resulting in the need for 3 additional procedures to take place at later, unspecified, dates to drain the excess pleural fluid (patient id (b)(6)).Investigation ¿ evaluation: reviews of documentation including the review of complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures, as well as a visual inspection of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.A database search did not identify any other complaints associated with the reported device lot.The information provided upon review of complaint file, device history record, complaint history, and quality control documents did not provide evidence to support that the device was manufactured out of specification or to suggest items in the lot or similar devices in the field or in house were nonconforming.Cook also reviewed product labeling.The product ifu, [tfbfbp_rev1] ¿harrison fetal bladder stent set¿ contains the following information related to the reported failure mode.¿assembly instructions 1.Just prior to the placement procedure, load the positioner onto the wire guide.2.Load the stent onto the wire guide by introducing the wire guide into the multi-length coil of the stent until 3mm- 4mm of the wire guide extends beyond the distal coil of the stent.Place non-flared end of the positioner against the end of the multi-length (proximal) stent coil¿¿ ¿implanting harrison fetal bladder stent.¿ 4.Trocar tip should be advanced 5mm -10cm into the fetal bladder.¿ 5.While stabilizing the needle, advance the stent assembly into the needle.6.After the positioner has entered the hub of the needle and is within the needle cannula, stabilize the positioner and remove the wire guide.¿ based on the available information, no product returned, and the results of the investigation, the cause of the incident was unable to be determined.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|