MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87215
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A licensed practical nurse (lpn) from a long-term care (ltc) facility reported to fresenius technical support (ts) that a fluid leak occurred during a patient's peritoneal dialysis (pd) treatment.Prior to discovery of the fluid leak, it was reported that an ¿m31 air detected in cassette¿ alarm occurred during drain 3 of 6.The lpn contacted ts for assistance in bypassing the alarm.After failing to move past the alarm, the lpn was advised to resetup the cycler with new supplies.When the lpn opened the cycler door, the fluid leak was discovered.The lpn confirmed there was no visible damage on the back of the cassette.The origin of the leak was not known.The patient did not complete their treatment.The lpn said the staff at the facility were never trained to perform manual pd exchanges as an alternative to using the cycler.The lpn was advised by ts to let the cycler dry out and wait a full 24-hours before using it again.In order to do this, the patient had to miss treatment the following day as well.However, the lpn confirmed the patient did not develop any symptoms, experience any adverse effects, injuries, or require medical intervention as a result of the reported event.The lpn confirmed the patient¿s doctor was notified, and prophylactic antibiotics were not prescribed.The patient was said to be continuing pd therapy on the same cycler without any further issues.The cycler set was not available to be returned for manufacturer evaluation as it had been discarded.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
A licensed practical nurse (lpn) from a long-term care (ltc) facility reported to fresenius technical support (ts) that a fluid leak occurred during a patient's peritoneal dialysis (pd) treatment.Prior to discovery of the fluid leak, it was reported that an ¿m31 air detected in cassette¿ alarm occurred during drain 3 of 6.The lpn contacted ts for assistance in bypassing the alarm.After failing to move past the alarm, the lpn was advised to resetup the cycler with new supplies.When the lpn opened the cycler door, the fluid leak was discovered.The lpn confirmed there was no visible damage on the back of the cassette.The origin of the leak was not known.The patient did not complete their treatment.The lpn said the staff at the facility were never trained to perform manual pd exchanges as an alternative to using the cycler.The lpn was advised by ts to let the cycler dry out and wait a full 24-hours before using it again.In order to do this, the patient had to miss treatment the following day as well.However, the lpn confirmed the patient did not develop any symptoms, experience any adverse effects, injuries, or require medical intervention as a result of the reported event.The lpn confirmed the patient¿s doctor was notified, and prophylactic antibiotics were not prescribed.The patient was said to be continuing pd therapy on the same cycler without any further issues.The cycler set was not available to be returned for manufacturer evaluation as it had been discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE PATIENT CONN
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key17715830
Report Number0008030665-2023-00852
Device Sequence Number15140875
Product Code FKX
UDI-Device Identifier00840861100743
UDI-Public00840861100743
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K173718
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue Number050-87215
Device Lot Number22LR08065
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2023
Supplement Date Received by Manufacturer11/09/2023
Initial Report FDA Received Date09/08/2023
Supplement Report FDA Received Date11/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID.; DELFLEX PD FLUID.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.
Patient Age57 YR
Patient SexFemale
Patient Weight126 KG
-
-