|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct h8.Three (3) good faith effort (gfe) attempts were made to retrieve the reporter's title and occupation and a response from the customer was not provided.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is possible that the tissue pad fell off due to the following mechanism.1.Part of the tissue pad peeled off because the seal & cut output was performed when nothing was gripped in the grip (including after the tissue was cut).2.The tissue pad fell off because a load was applied to the peeled off tissue pad.The event can be detected/prevented by following the instructions for use which state: "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation".Olympus will continue to monitor field performance for this device.
|