• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES (IRELAND) LTD. DUO HEADLIGHT 2 BAY SYSTEM - US; PFM05

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES (IRELAND) LTD. DUO HEADLIGHT 2 BAY SYSTEM - US; PFM05 Back to Search Results
Catalog Number 90620US
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The duo headlight 2 bay system - us (90620us) was returned for evaluation: the duo headlight 2 bay system was received in used condition.The depot technician completed the evaluation of the duo headlight and could not duplicate any issues with the duo headlight overheating or flickering.The unit passed all tests.The customer reported that the fan stops intermittently.It was determined that the issue could be a malfunctioning battery holster issue; however, the customer did not return the unit with the holster for evaluation.The unit will be send back with a new holster to correct the fan stop issue.The reported complaint could not be confirmed.No manufacturing, workmanship, or material deficiency has been identified.
 
Event or Problem Description
It was reported that the cooling fan of the duo headlight 2 bay system - us (90620us) operated intermittently, causing the light to overheat and flicker.It is unknown under what circumstance this event occurred; however, no patient injury, death or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO HEADLIGHT 2 BAY SYSTEM - US
Common Device Name
PFM05
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17716543
Report Number3006697299-2023-00099
Device Sequence Number6891149
Product Code FSR
UDI-Device Identifier10381780490746
UDI-Public10381780490746
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number90620US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2023
Initial Report FDA Received Date09/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-