| Model Number |
A37025A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The evaluation found a broken eyepiece.Furthermore, the following additional issues were identified during inspection: outer tube bending, peeling of gold plating on outer tube, and bad illumination due to broken fibers.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus, the op telescope, had a cracked eyepiece.The issue was found during inspection for use.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|