• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC PERFORM REV INSERT SZ 3/4, 42MM DIA, +0; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC PERFORM REV INSERT SZ 3/4, 42MM DIA, +0; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number DWP2420
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Joint Laxity (4526); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
There was a revision of a tornier product, specifically a perform reverse glenoid with a perform humeral stem.The baseplate and stem were not removed but were adjusted by lateralizing the glenoid sphere.A spacer and a new poly component were used on the humeral side.The revision was necessary due to inadequate tensioning in the initial procedure.An update confirmed that the revision was successful, and it involved changing a standard glenosphere for a lateralized one and building up the poly component on the humeral side.
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device disposition is unknown.
 
Event or Problem Description
There was a revision of a tornier product, specifically a perform reverse glenoid with a perform humeral stem.The baseplate and stem were not removed but were adjusted by lateralizing the glenoid sphere.A spacer and a new poly component were used on the humeral side.The revision was necessary due to inadequate tensioning in the initial procedure.An update confirmed that the revision was successful, and it involved changing a standard glenosphere for a lateralized one and building up the poly component on the humeral side.
 
Additional Manufacturer Narrative
Please note the corrections made to the device code, clinical signs code, and health impact code: the reported event was not confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM REV INSERT SZ 3/4, 42MM DIA, +0
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17718466
Report Number0001649390-2023-00228
Device Sequence Number7176661
Product Code PHX
UDI-Device Identifier00846832084517
UDI-Public00846832084517
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K201315
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/12/2026
Device Catalogue NumberDWP2420
Device Lot Number2801AX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2023
Supplement Date Received by Manufacturer10/11/2023
Initial Report FDA Received Date09/11/2023
Supplement Report FDA Received Date11/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-