| Model Number |
02-01609 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Superficial (First Degree) Burn (2685)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient reported that they went to urgent care for first degree burns under the patch.The patient was prescribed ointment and switched to flex adaptor and continued with service.After swithcing to flex the patient returned to urgent care and received more ointment due to first degree burns.The patient returned the kit and did not finish service.The patient does have a history of skin sensitivity.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is related to report number 2133409-2023-00050.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that patient experienced a first degree burn under the universal patch.The universal patch was not returned for investigation.Engineering evaluation was unable to be performed as the electrode/device was not returned.Allegation is confirmed through the need for a prescription and is most probable to be a bio-incompatibility issue with the electrode adhesive.Marsi, skin burn, and associated symptoms may inherently occur under the course of ecg monitoring.No single factor or combination of factors can be attributable to electrode skin irritation and associated symptoms.The product labeling advises patients of alternate options and other steps to take if skin irritation develops, including healthcare professional contact as needed.
|
| |
|