• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BRAEMAR MANUFACTURING, LLC C6 MCOT PPM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BRAEMAR MANUFACTURING, LLC C6 MCOT PPM Back to Search Results
Model Number 02-01609
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Superficial (First Degree) Burn (2685)
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported that they went to urgent care for first degree burns under the patch.The patient was prescribed ointment and switched to flex adaptor and continued with service.After swithcing to flex the patient returned to urgent care and received more ointment due to first degree burns.The patient returned the kit and did not finish service.The patient does have a history of skin sensitivity.
 
Additional Manufacturer Narrative
This report is related to report number 2133409-2023-00050.
 
Additional Manufacturer Narrative
It was reported that patient experienced a first degree burn under the universal patch.The universal patch was not returned for investigation.Engineering evaluation was unable to be performed as the electrode/device was not returned.Allegation is confirmed through the need for a prescription and is most probable to be a bio-incompatibility issue with the electrode adhesive.Marsi, skin burn, and associated symptoms may inherently occur under the course of ecg monitoring.No single factor or combination of factors can be attributable to electrode skin irritation and associated symptoms.The product labeling advises patients of alternate options and other steps to take if skin irritation develops, including healthcare professional contact as needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C6 MCOT PPM
Common Device Name
C6 MCOT PPM
Manufacturer (Section D)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer (Section G)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer Contact
beverly okoh
1285 corporate center d
suite 150
eagan, MN 55121
MDR Report Key17718911
Report Number2133409-2023-00049
Device Sequence Number7073082
Product Code DSI
UDI-Device IdentifierB146PE20
UDI-PublicB146PE20
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K081444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number02-01609
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2023
Supplement Date Received by Manufacturer01/24/2024
Initial Report FDA Received Date09/11/2023
Supplement Report FDA Received Date02/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ELECTRODE - S&W, 470-0021-01
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-