ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 48C; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 00625006535, item name: bone scr 6.5x35 self-tap.Lot # j7410192.G2: foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the cup was implanted and needed a screw for better fixation, the surgeon thinks he over tightened the screw causing it to go through the cup.Cup and screw and liner was removed after intra op x-ray was preformed and replaced with identical implants.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: limited evaluation of a right total hip arthroplasty with single screw which has penetrated through the implant and implanted fully into bone.It was stated the surgeon thinks he may have over tightened the screw but a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 48C
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17719905
Report Number0001825034-2023-02136
Device Sequence Number7213359
Product Code PBI
UDI-Device Identifier00880304524200
UDI-Public(01)00880304524200(17)311207(10)7151430
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000661
Device Lot Number7151430
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2023
Supplement Date Received by Manufacturer12/20/2023
Initial Report FDA Received Date09/11/2023
Supplement Report FDA Received Date01/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE H10.
Patient SexPrefer Not To Disclose
-
-