• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND DIAGNOSTICS DIVISION MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND DIAGNOSTICS DIVISION MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Catalog Number 08P19-34
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed falsely elevated magnesium results on alinity c processing module for one patient.The results provided were: on (b)(6) 2023, sid (b)(6) initial=3.29 mmol/l /repeated=0.70 and 0.69 mmol/l there was no reported impact to patient management.
 
Event or Problem Description
The customer observed falsely elevated magnesium results on alinity c processing module for one patient.The results provided were: on (b)(6)2023 sid (b)(6)initial=3.29 mmol/l /repeated=0.70 and 0.69 mmol/l there was no reported impact to patient management.
 
Additional Manufacturer Narrative
The complaint evaluation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history records for alinity c magnesium reagent, lot number 52268ud00.The ticket search determined that there is as expected complaint activity for the likely cause lot.Trending review determined no trend for the issue for the product.Return testing was not completed as returns were not available.Device history record review did not identify any non-conformances or deviations with the complaint lot and complaint issue.Historical performance of the alinity c magnesium reagents in the field was reviewed using data gathered via abbottlink from customers worldwide.The median value of the patient population for the complaint lot 52268ud00 is within the established limits and comparable to historical reagent lot performance.All field data demonstrating that the lot is performing as expected.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the alinity c magnesium reagent, lot number 52268ud00 was identified.Updated information in section g1: contact office address 1, contact office city, contact office postal code, contact office country, contact office first and last name, contact office email, contact office phone number, contact office fax number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key17720940
Report Number3005094123-2023-00236
Device Sequence Number12566015
Product Code JGJ
UDI-Device Identifier00380740176532
UDI-Public00380740176532
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2024
Device Catalogue Number08P19-34
Device Lot Number52268UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/05/2023
Supplement Date Received by Manufacturer12/08/2023
Initial Report FDA Received Date09/11/2023
Supplement Report FDA Received Date12/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/09/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, (B)(6); ALNTY C PROCESSING MODU, 03R67-01, (B)(6)
Patient SexUnknown
-
-