ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER; STOPCOCK, I.V. SET Back to Search Results
Catalog Number B30183
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complete d1 - brand name is 32" (81 cm) appx 6.3 ml, 15 drop admin set w/0.2 micron filter, rotating luer w/filter cap, bag hanger.The device was returned for evaluation.The investigation is pending.
 
Event or Problem Description
The event involved a 32" (81 cm) appx 6.3 ml, 15 drop admin set w/0.2 micron filter, rotating luer w/filter cap, bag hanger where an air in line towards the end of durvalumab infusion was reported at 12:59.The infusion rate was 280 ml/hr.There was patient involvement and unknown patient harm reported.
 
Additional Manufacturer Narrative
One used sample list #b30183 connected with spiros, an used bd phaseal and a fresenius kabi 0.9% sodium chloride injection usp freeflex 500 ml intravenous bag with durvalumab were returned for evaluation.The inlet and outlet filters appeared to be wetted out.No additional damage or anomaly was observed.The set was primed and purged as per procedure and a leaks coming from the outlet filter vent was confirmed.No additional leaks or air in line were observed.Complaint of product incorporate / allows air passage cannot be confirmed, due to no air after purge the line was confirmed.Small bubbles trapped inside the filter were observed, however, that's an indication of normal functionality of the filter.Only a leak coming from outlet filters was confirmed.The probable cause is typical due to a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key17722844
Report Number9617594-2023-00680
Device Sequence Number17946787
Product Code FMG
UDI-Device Identifier00887709055445
UDI-Public(01)00887709055445(17)270901(10)9930806
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K964435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberB30183
Device Lot Number9930806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2023
Supplement Date Received by Manufacturer09/20/2023
Initial Report FDA Received Date09/11/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DURVALUMAB, MFR UNK.
Patient SexUnknown
-
-