Catalog Number B2050-040
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Codes Pain (1994); Embolism/Embolus (4438)
Date of Event 02/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified femoral artery.The 5x40mm armada 35 balloon dilatation catheter (bdc) was advanced to the lesion.The balloon was over inflated and then fully deflated.During removal, resistance was noted with the anatomy, extra force was used and the bdc separated into four parts.Only three parts were removed from the patient via snare.The patient experienced pain and it was noted that an embolism had occurred in one of the axis of the leg.The fourth piece was removed via surgery a few weeks later.There was no adverse patient sequela; however, clinically significant delay was noted during the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.B3: date of event estimated.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It was reported that the balloon was overinflated during the procedure.It should be noted that the percutaneous transluminal angioplasty catheter (pta), armada 35/ armada 35ll global, ce, instruction for use (ifu) states: ¿inflation in excess of the rated burst pressure may cause the balloon to rupture.Use of a pressure monitoring device is recommended.¿ the rated burst pressure for this device is 22 atmospheres as indicated on the product label.In this case since it is unknown to what atmosphere the balloon was overinflated, it is unknown if the ifu violation contributed to the reported complaint.The investigation determined the reported difficulty removing the device, balloon separation, delay to treatment/therapy, hospitalization or prolonged hospitalization, surgical intervention and unexpected medical intervention appear to be related to operational context.A conclusive cause for the reported patient effects of embolism and pain and the relationship to the device, if any, cannot be determined.The reported patient effects of embolism and pain are listed in the percutaneous transluminal angioplasty (pta) catheter, armada 35 / armada 35 ll, global ifu as known patient effects.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Event or Problem Description
Subsequently, after the initial report was filed it was noted the fourth piece was removed via open surgery.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17722858
Report Number2024168-2023-09883
Device Sequence Number7219682
Product Code LIT
UDI-Device Identifier08717648155178
UDI-Public08717648155178
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Catalogue NumberB2050-040
Device Lot Number20216G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2023
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date09/11/2023
Supplement Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient SexUnknown
-
-