| Catalog Number |
72203853 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
08/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, after the sututerfix ultra was placed bone hole; when the anchor was being tied, it came out from the glenoid.The procedure was completed with non-significant surgical delay using a smith and nephew back up device in an additional drilled bone hole.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that the photos of the device and packaging box was provided for review however the photos do not indicated a clinical root cause for the reported event.Based on the limited information provided the clinical root cause of the reported events cannot be determined and we are currently unable to rule out a procedural variance as a contributing factor to the reported event, which does not represent a device malfunction.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|