• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC WEBSTER¿ ELECTROPHYSIOLOGY CATHETER; CATHETER, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC WEBSTER¿ ELECTROPHYSIOLOGY CATHETER; CATHETER, ELECTRODE RECORDING Back to Search Results
Catalog Number D610DRP10RT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6.Investigation findings code of "appropriate term/code not available" represents photo/video analysis.A picture was received for evaluation following biosense webster's procedures.According to pictures provided by the customer, the tip was observed broken without sharp edges or cable exposed.A manufacturing record evaluation was performed for the finished device 30925052m number, and no internal action related to the complaint was found during the review.The customer complaint was confirmed from the photo analysis.However, the device has not been returned for evaluation.Since no device has been received, no product investigation can be performed.If the device is received in the future, the product analysis will be performed as appropriate in order to find the root cause of the complaint.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a supraventricular tachycardia (svt) cardiac ablation procedure with a webster¿ electrophysiology catheter and during the operation, the tip or splines of the catheter was found bent (no exposed wires).A second device was used to complete the operation.There was no adverse event reported on patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEBSTER¿ ELECTROPHYSIOLOGY CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17726512
Report Number2029046-2023-02048
Device Sequence Number5820416
Product Code DRF
UDI-Device Identifier10846835000238
UDI-Public10846835000238
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K892265
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2025
Device Catalogue NumberD610DRP10RT
Device Lot Number30925052M
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2023
Initial Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-