The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the balloon feeder return tube clogged.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the balloon feeder return tube.In addition, our technician confirmed that the segment steel braid broken, the light guide cable buckled, the warning/caution label worn out, the ultrasound connector body defective, the insertion flexible tube bump, the light guide cable bump, the us connector cable (long) bump, and the us probe eus image no scanline; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0630(air/water & jet water channels)"" and/or the risk analysis results, it was evaluated to submit mdr.
|