Customer reported issue with introducer having fluid coming back up into the distal side port, air bubbles coming up the proximal line, and fluid leaking from the juncture hub at the skin.It was reported the catheters work great in the or and then 1-9 days later experience the issue.The introducer was removed and a different catheter was placed.No patient injury reported.
|
|
Qn# (b)(4).As per the medwatch report received on 14sep2023, rij 9 fr double lumen with pa us guided placed and removed 11 days later.Patient had a rij double cordis with a swan-ganz catheter.Patient was on continuous renal replacement therapy (crrt), so a calcium chloride, epinephrine, norepinephrine, and vasopressors all via the rij white lumen.Clinical nurse specialist (cns) has line.The customer report of a catheter leak could not be confirmed through investigation of the returned sample.The customer provided seven photos for analysis; and returned one mac for analysis.Definite signs of use in the form of biomaterial were observed on the catheter.Visual analysis of the catheter did not reveal any obvious defects or anomalies.The outer diameter (od) of the distal extension line measured 4.787mm via calibrated micrometer which was within the specification limits of 4.70mm-4.80mm per the distal extension line extrusion product drawing.The inner diameter (id) of the distal extension line measured 0.118" via calibrated pin gauge, or 2.9972mm, which was within the specification limits of 2.90mm-3.00mm per the distal extension line extrusion product drawing.The od of the proximal extension line measured 2.922mm via calibrated micrometer, which was within the specification limits of 2.90mm-3.00mm per the proximal extension line extrusion product drawing.The id of the proximal extension line measured 0.087" via calibrated pin gauge, or 2.2098mm, which was within the specification limits of 2.11mm-2.21mm per the proximal extension line extrusion product drawing.The valve and mac device were leak tested according to three different parameters.Low pressure leak resistance test states, "using a test pressure of 38-42 kpa (5.51-6.09 psi), there shall be no leakage past the hemostasis valve".The mac catheter was attached to the lab leak tester.With the distal end of the sheath occluded, the sheath was pressurized to 42 kpa for 30 seconds.No leaking or crossover was observed.Liquid leakage - hemostasis valve with mac companion inserted.The parameters from the low-pressure leak resistance test were repeated with a lab inventory 7fr mac companion inserted into the sheath.The sheath was pressurized to 42 kpa for 30 seconds and no leaks were observed from the hemostasis valve.High pressure leak resistance test states, "when tested in accordance with iso, using a test pressure of 300-320 kpa (43.5-46.4 psi) there shall be no leakage sufficient to form a falling drop." with both the distal end of the sheath and the hemostasis valve occluded with the returned obturator, the device was pressurized to 300kpa for 30 seconds.The extension lines were connected individually to the leak tester, and when pressurized, no catheter backflow or leaks were detected from any portion of the mac, which indicated that the assembly was intact.A manual tug test confirmed that the extension lines were secure within their respective luer hubs.The ifu provided with the kit informs the user, "do not suture directly to the outside diameter of access device to minimize the risk of cutting or damaging device or impeding device flow".A device history record review was performed, and no relevant findings were identified.Based on the customer report and the sample received, no problem was found.Teleflex will continue to monitor and trend on reports of this nature.
|
|
Customer reported issue with introducer having fluid coming back up into the distal side port, air bubbles coming up the proximal line, and fluid leaking from the juncture hub at the skin.It was reported the catheters work great in the or and then 1-9 days later experience the issue.The introducer was removed and a different catheter was placed.No patient injury reported.
|