| Catalog Number |
145FS35 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that purple coating color of the guidewire was peeling and could be seen under endoscopy.Stated that the user was exposed to hazardous, blood, or bodily fluids.The user needed to reinsert a new guidewire.It was unknown what medical intervention was provided.Per sample evaluation notification received on (b)(6) 2023, stated that the guidewire was found to be exposed during the sample evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed and the cause was unknown.A potential root cause for this failure is "improper material selection/geometry".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract." "do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract." "do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval." "failure to comply with the warnings could result in damage to the guidewire to include, but not limited to wire breakage, abrasion of the coating, release of guidewire fragments into the urinary system, all of which might require intervention." "should be used only by physicians thoroughly trained in endourological guidewire techniques." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that purple coating color of the guidewire was peeling and could be seen under endoscopy.Stated that the user was exposed to hazardous, blood, or bodily fluids.The user needed to reinsert a new guidewire.It was unknown what medical intervention was provided.Per sample evaluation notification received on 22aug2023, stated that the guidewire was found to be exposed during the sample evaluation.
|
| |
|