|
|
| Catalog Number |
ECR45M |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/12/2023.D4: batch # 879a90.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch history records were reviewed and certified by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unk surgery, after fired, noted the staples malformed.Changed to a new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/17/2023.D4: batch # 720a17.Investigation summary : the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload. visual analysis of the returned sample determined that one ecr45m reload was received unfired.The returned reload was loaded into a test device.The device was tested for functionality in the straight position and achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: tissue thickness should be carefully evaluated before firing any stapler.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.When positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc.Are within the instrument jaws.The event described could not be confirmed as the reload performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number 720a17, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|