| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty, the tip of the locking screw broke when the surgeon tightened the fixation screw of the articular surface with the torque wrench.He was not able to remove the broken part so he decided to implant fixed ps articular surface instead.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - japan.The device will not be returned for analysis as it remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified a screw that was fractured and missing the threads.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|